omnipod insulin_management_system_firmware 访问控制不恰当

admin 2023-11-30 15:53:09 Ali_nvd 来源:ZONE.CI 全球网 0 阅读模式
omnipod insulin_management_system_firmware 访问控制不恰当

CVE编号

CVE-2020-10627

利用情况

暂无

补丁情况

N/A

披露时间

2021-12-02
漏洞描述
Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
解决建议
建议您更新当前系统或软件至最新版,完成漏洞的修复。
参考链接
https://us-cert.cisa.gov/ics/advisories/icsma-20-079-01
https://www.myomnipod.com/security-bulletins
受影响软件情况
# 类型 厂商 产品 版本 影响面
1
运行在以下环境
系统 omnipod insulin_management_system_firmware - -
运行在以下环境
硬件 omnipod insulin_management_system 19191 -
运行在以下环境
硬件 omnipod insulin_management_system 40160 -
CVSS3评分 8.1
  • 攻击路径 相邻
  • 攻击复杂度 低
  • 权限要求 无
  • 影响范围 未更改
  • 用户交互 无
  • 可用性 无
  • 保密性 高
  • 完整性 高
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
CWE-ID 漏洞类型
CWE-284 访问控制不恰当
NVD-CWE-Other
- avd.aliyun.com
weinxin
版权声明
本站原创文章转载请注明文章出处及链接,谢谢合作!
N/A Ali_nvd

N/A

N/ACVE编号 CVE-2024-9120利用情况 暂无补丁情况 N/A披露时间 2024-09-23漏洞描述Use after free in Dawn