omnipod insulin_management_system_firmware 访问控制不恰当
CVE编号
CVE-2020-10627利用情况
暂无补丁情况
N/A披露时间
2021-12-02漏洞描述
Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.解决建议
建议您更新当前系统或软件至最新版,完成漏洞的修复。
参考链接 |
|
---|---|
https://us-cert.cisa.gov/ics/advisories/icsma-20-079-01 | |
https://www.myomnipod.com/security-bulletins |
受影响软件情况
# | 类型 | 厂商 | 产品 | 版本 | 影响面 | ||||
1 | |||||||||
---|---|---|---|---|---|---|---|---|---|
运行在以下环境 | |||||||||
系统 | omnipod | insulin_management_system_firmware | - | - | |||||
运行在以下环境 | |||||||||
硬件 | omnipod | insulin_management_system | 19191 | - | |||||
运行在以下环境 | |||||||||
硬件 | omnipod | insulin_management_system | 40160 | - |
- 攻击路径 相邻
- 攻击复杂度 低
- 权限要求 无
- 影响范围 未更改
- 用户交互 无
- 可用性 无
- 保密性 高
- 完整性 高
CWE-ID | 漏洞类型 |
CWE-284 | 访问控制不恰当 |
NVD-CWE-Other |